Harvard Tracer Suite (HTS)

HTS is composed of two instruments based on absorption of near-infrared laser radiation in high finesse optical cavities. A Picarro G2401-m analyzer based on wavelength-scanned cavity ring-down spectroscopy (CRDS) measures CO2, CH4, and CO concentrations at 2-second intervals. A Los Gatos 913-0014 EP analyzer based on off-axis integrated cavity output spectroscopy (ICOS) measures N2O and CO concentrations at 1-second intervals. Extensive modifications have been applied to these commercial analyzers for flight and include vibration isolation, temperature control, additional flow control and pumping capacity for high-altitude sampling, sample drying, and in-flight calibrations using WMO-traceable compressed gas standards to verify stable and accurate performance throughout the full DC-8 flight envelope.