The far-infrared spectroscopy of the troposphere (FIRST) instrument is a Fourier-transform spectrometer developed to measure the Earth’s thermal emission spectrum with a particular emphasis on the far-infrared. FIRST has observed the atmosphere from both the ground looking up and from a highaltitude balloon looking down. A recent absolute laboratory calibration of FIRST under ground-like operating conditions showed accuracy to better than 0.3 K at near-ambient temperatures (270–325 K) but reduced accuracy at lower temperatures. This paper presents calibration results for balloon-flight conditions using a cold blackbody to simulate the space view used for on-board calibration. An unusual detector nonlinearity was discovered and corrected, and stray light was measured and removed. Over most of the range of Earth scene temperatures (205–300 K), the accuracy of FIRST is 0.35–0.15 K (one sigma).
Far-infrared spectroscopy of the troposphere: calibration with a cold background
Latvakoski, H., M. Mlynczak, R. Cageao, D. Johnson, and D. Kratz (2014), Far-infrared spectroscopy of the troposphere: calibration with a cold background, Appl. Opt., 53, 5425-5433.
Abstract
Research Program
Radiation Science Program (RSP)
Disclaimer: This material is being kept online for historical purposes. Though accurate at the time of publication, it is no longer being updated. The page may contain broken links or outdated information, and parts may not function in current web browsers. Visit https://espo.nasa.gov for information about our current projects.