The Picarro CO2/CH4 Flight Analyzer is a real time, trace gas monitor capable of measuring these gases with parts-per-billion (ppbv) sensitivity onboard aircraft with varying cabin pressure and environmental conditions. The analyzer is based on Wavelength-Scanned Cavity Ring Down Spectroscopy (WS-CRDS), a time-based measurement utilizing a near-infrared laser to measure a spectral signature of the molecule. Gas is circulated in an optical measurement cavity with an effective path length of up to 20 kilometers. A patented, high-precision wavelength monitor makes certain that only the spectral feature of interest is being monitored, greatly reducing the analyzer’s sensitivity to interfering gas species, and enabling ultra-trace gas concentration measurements even if there are other gases present. As a result, the analyzer maintains high linearity, precision, and accuracy over changing environmental conditions with minimal calibration required.
Picarro G1301-c Methane/Carbon Dioxide Analyzer
Instrument Type
Aircraft
Point(s) of Contact
(POC; PI),
(PI)
Disclaimer: This material is being kept online for historical purposes. Though accurate at the time of publication, it is no longer being updated. The page may contain broken links or outdated information, and parts may not function in current web browsers. Visit https://espo.nasa.gov for information about our current projects.