CU Aircraft Extractive Electrospray Time-of-Flight Mass Spectrometer

Status

Status
Operational
(Currently fully integrated with CU HR-AMS)
Operated By
PI

Principle: The CU aircraft Extractive Electrospray Time-of-Flight Mass Spectrometer (EESI-TOFMS) detects the chemical composition of submicron particulate matter by simultaneous dissolution of the aerosol and soft ionization of its molecular components in an electrospray, followed by detection using time-of-flight mass spectrometry. When operated with positive ion polarity analytes are detected intact as adducts with sodium ions. When operated with negative ion polarity analytes are detected as deprotonated anions.

 

Aircraft Operation: Operated with positive or negative ion polarity, depending on mission goals.
Pressure-controlled EESI region with a maximum sampling altitude of 23 kft. Sensitivity can be increased for boundary layer  sampling by increasing the pressure of the electrospray region.
Calibrated relative to AMS on flight days, and confirmed with absolute calibrations on maintenance days
1s time resolution with 15s background measurements every 4 min.

 

Data Products:
Positive ion polarity EESI(+): C6H10O5 calibrated as levoglucosan and C6H5NO4 calibrated as nitrocatechol
Negative ion polarity EESI(-): C6H5NO4 calibrated as nitrocatechol
Additional species under development. Species of interest for specific campaigns can be tested and calibrated for.

 

Detection Limits (1s):
C6H10O5 as levoglucosan: 200 ng sm-3
C6H5NO4 as nitrocatechol: 20ng sm-3 EESI(-), 400 ng sm-3 EESI(+)

Instrument Type
Measurements
Aircraft
Recent Missions
(DC-8 - AFRC)
Point(s) of Contact
Range of Measurement
In situ
Measurement Sampling Rate
1.00 Hz
Data Delivery
24 h for preliminary data, 6-9 months for final data
Weight
210.00 kg
Size
104.00 cm (L) x 61.00 cm (W) x 135.00 cm (H)
Power:
450.00 W
(maximum
1000.00 W
)
Has Pressurized Canisters? Yes
Location
Dual rack installation inside the cabin. Inlet mounted on a custom window plate (DC-8) or on a belly port (NSF C-130)
Notes
Installation weights and measures refer to the single bay installation during FIREX-AQ in combination with the CU HR-AMS. An additional calibration rack (95x61x60 cm, 92 kg) was flown and is needed for proper airborne instrument operation.
TRL
9
Publications

 

Disclaimer: This material is being kept online for historical purposes. Though accurate at the time of publication, it is no longer being updated. The page may contain broken links or outdated information, and parts may not function in current web browsers. Visit https://espo.nasa.gov for information about our current projects.