CSU QC-TILDAS Ammonia

Status

Status
Operational
(Currently in preparation for the AEROMMA mission on the DC-8)
Operated By
PI

Ambient ammonia (NH3) mixing ratios are measured in-situ using a flight-ready, closed-path, optical-based NH3 monitoring system. The CSU-NH3 instrument system consists of a combination of commercially-available and custom-built components including: 1) a commercially-available infrared absorption spectrometer that serves as the heart of the NH3 monitor, 2) a commercially-available inertial inlet that acts as a filter-less separator of particles from the sample stream, 3) a custom-built aircraft inlet, 4) a custom-designed vibration isolation mounting system for the spectrometer, and 5) an optional system for adding passivant to the sample stream.

The heart of the instrument is a closed-path, commercial (Aerodyne Research, Inc.), single-channel, quantum-cascade tunable infrared laser direct absorption spectrometer (QC-TILDAS) [McManus et al., 2010; McManus et al., 1995; Zahniser et al., 1995]. This spectrometer uses a direct absorption technique combined with a high sample flow rate (>10 SLPM) to achieve fast (up to 10 Hz) collection of absolute NH3 mixing ratios. The QC-TILDAS is operated with a heated aerodynamic separator (Aerodyne Research Inc., Inertial Inlet) that provides filter-less separation of particles >300 nm from the sample stream [Ellis et al., 2010]. An injection-style aircraft inlet allows calibration gases to be introduced into the sample stream within a few centimeters of the inlet tip. The custom inlet system is also designed to support the option for active continuous passivation of the sampling sufaces by 1H,1H-perflurooctylamine, a strong perfluorinated base that acts to coat the sampling surfaces with nonpolar chemical groups. Injection of this chemical into the aircraft inlet near the inlet tip prevents adsorption of both water and basic species on the sampling surfaces. The coating has been shown to greatly improve the instrument's time response in the laboratory and aboard research aircraft by increasing transmission of NH3 through the sample flow path [Pollack et al., 2019; Roscioli et al., 2016].

The QC-TILDAS is regularly calibrated on the ground and in flight via standard addition to the sample stream with a known concentration of NH3 generated from a temperature-regulated permeation tube (Kin-Tech), and zeroed by overflowing the inlet tip with a bottled source of NH3-free, synthetic air. The emission rate of the permeation device is calibrated before and after every mission by the NOAA ultraviolet optical absorption system [Neuman et al., 2003]. Allan variance analyses indicate that the in-flight precision of the instrument is 60 ppt at 1 Hz corresponding to a 3-sigma detection limit of 180 ppt. Zero signals span ±200 pptv, or 400 pptv total, with fluctuations in cabin pressure and temperature and altitude in flight. The total uncertainty associated with the 1-Hz measurement is ±(12% of the measured mixing ratio + 200 pptv).

The CSU-NH3 instrument has been successfully deployed (i.e. 100% data coverage) in two prior airborne research campaigns; one on the NSF/NCAR C-130 aircraft during the 2018 Western wildfire Experiment of Cloud Chemistry, Aerosol absorption and Nitrogen (WE-CAN) field campaign and the other aboard the University of Wyoming King Air during the TRANS2Am field campaign in 2019, 2021, and 2022. The aircraft inlet and aerodynamic separator are currently being modified in the laboratory to support lower pressure altitudes such as those anticipated for the full altitude range of the NASA DC-8 aircraft.

Instrument Type
Measurements
Recent Missions
(DC-8 - AFRC)
;
(University of Wyoming King Air)
;
(NSF/NCAR C-130)
Point(s) of Contact
(POC; Co-I)
Range of Measurement
In situ
Measurement Sampling Rate
1.00 Hz
Data Delivery
24 hours for in-field data; 3-6 weeks to produce final data
Has Pressurized Canisters? Yes
Location
The instrument system described above in the configuration that it was utilized aboard the C-130 aircraft during WE-CAN requires the space of an entire NSF/NCAR G-V aircraft equipment rack (approximate dimensions 21.5” W x 28” D x 50” H). The equipment without the rack weighed approximately 150 kg and included a 30 kg uninterruptable power supply (UPS) and a 10 kg display laptop. The total power used by the instrument system was 1600 watts, with roughly one third of this total (600 watts) being dedicated to the main pumping system (Agilent, model Triscroll 600, 100 lbs installed). It is possible that the power, weight and space required for this instrument system can be reduced for future deployments by eliminating the UPS and display laptop. It may also be possible to reduce the size of the pump if different field applications allow for a lower sample flow rate to be used. A size 150A cylinder of compressed UHP N2 is used with the instrument in flight. The instrument is currently being outfitted for a single bay of a DC-8 high rack in preparation for the 2023 AEROMMA mission.

 

Disclaimer: This material is being kept online for historical purposes. Though accurate at the time of publication, it is no longer being updated. The page may contain broken links or outdated information, and parts may not function in current web browsers. Visit https://espo.nasa.gov for information about our current projects.