Chemical Ionization Mass Spectrometer

Status

Status
Operational
Operated By
PI

The NOAA chemical ionization mass spectrometer (CIMS) instrument was developed for high-precision measurements of gaseous nitric acid (HNO3) specifically under high- and variable-humidity conditions in the boundary layer. The instrument’s background signals (i.e., signals detected when HNO3-free air is measured), which depend on the humidity and HNO3 concentration of the sample air, are the most important factor affecting the limit of detection (LOD). A new system to provide HNO3-free air without changing both the humidity and the pressure of the sampled air was developed to measure the background level accurately. The detection limit was about 23 parts per trillion by volume (pptv) for 50-s averages. Field tests, including an intercomparison with the diffusion scrubber technique, were carried out at a surface site in Tokyo, Japan, in October 2003 and June 2004. A comparison between the measured concentrations of HNO3 and particulate nitrate indicated that the interference from particulate nitrate was not detectable (i.e., less than about 1%). The intercomparison indicated that the two independent measurements of HNO3 agreed to within the combined uncertainties of these measurements.

Instrument Type
Measurements
Recent Missions
(DC-8 - AFRC)
;
(WB-57 - JSC)
;
(WB-57 - JSC)
;
(WB-57 - JSC)
;
(WB-57 - JSC)
;
(WB-57 - JSC)
Point(s) of Contact
Range of Measurement
In situ
Weight
214.00 kg
Size
Length
91.00 cm
(L) x
Width
76.00 cm
(W) x
Height
86.00 cm
(H)
Power:
2800.00 W
Has Pressurized Canisters? Yes
Location
Pallet

 

Disclaimer: This material is being kept online for historical purposes. Though accurate at the time of publication, it is no longer being updated. The page may contain broken links or outdated information, and parts may not function in current web browsers. Visit https://espo.nasa.gov for information about our current projects.