Disclaimer: This material is being kept online for historical purposes. Though accurate at the time of publication, it is no longer being updated. The page may contain broken links or outdated information, and parts may not function in current web browsers. Visit https://espo.nasa.gov for information about our current projects.

 

PeroxyAcetylNitrate, Aldehydes and Ketones (PANAK)

Status

Operated By: 
PI

Don Hearlth, Rick Kolyer

The Ames PANAK instrument is a computerized 3- channel Capillary Gas Chromatographic system designed for the collection and analysis of low ppt (10-12 v/v) levels of peroxyacyl nitrates (PANs), alkyl nitrates, and tertrachloroethene in Channels 1 and 2; and C2-C3 aldehydes, C1-C2 alcohols, C3-C4 ketones, and C1-C2 nitriles in channel 3. Channels 1 and 2 use ECD detectors and have a sampling frequency of 2.5 minutes. Channel 3 uses a Photo Ionization detector placed in series with a Reduction Gas detector and has a sampling frequency of 5 minutes. The main manifold draws 5 SL/min of ambient air through a heated Teflon lined probe from which each of the three instrument channels draws a 200 ml aliquot of sample air. This aliquot is dried by passing it through a –35 °C cold trap, cooled to -140 °C for constituent pre concentration, and then heat desorbed into the gas chromatographic columns. All calibrations are performed in-flight by using an installed dilution system and in a manner that mimics ambient air sampling. Primary standards are generally referred to a series of permeation tubes. In addition high concentration standards are also carried on board. Sensitivities under typical conditions are: 1-3 ppt PANs, 1-5 ppt alkyl nitrates, 5-20 ppt OVOC, and 20-30 ppt nitriles.

Instrument Type: 
Aircraft: 
Point(s) of Contact: 
Hanwant Singh (POC; PI)