UAS Chromatograph for Atmospheric Trace Species

Status

Status
Operational
Operated By
PI

The Unmanned Aircraft Systems (UAS) Chromatograph for Atmospheric Trace Species (UCATS) was designed and built for autonomous operation on remotely piloted aircraft, but has also been used on manned aircraft. It uses chromatography to separate atmospheric trace gases along narrow heated columns, followed by precise and accurate detection with electron capture detectors. There are currently three chromatography channels on UCATS, which measure nitrous oxide (N2O) and sulfur hexafluoride (SF6); CFC-11, CFC-12, CFC-113, and halon 1211; and chloroform (CHCl3) and carbon tetrachloride. On an earlier version of UCATS, with only two channels, we also measured methane, hydrogen, and carbon monoxide, along with N2O and SF6. In addition, there is a small ozone instrument and a tunable diode laser instrument for water vapor. Gas is pumped into the instruments from an inlet outside the aircraft, measured, and vented. UCATS has flown on the Altair UAS, the GV during HIPPO, the NASA Global Hawk UAS during the Global Hawk Pacific (GloPac) and ATTREX missions, where a record was set for the longest duration research flight (more than 28 hours), the DC-8 for ATom, and the ER-2 for DCOTSS. UCATS is relatively lightweight and compact, making it ideal for smaller platforms, but it is easily adaptable to a mid-size platform like the GV or Global Hawk. The data are used to measure sources and sinks of trace gases involved in climate and air quality, as well as transport through the atmosphere.

UCATS is three different instruments in one enclosure:

1. 3-channel (formerly 2-channel, up until 2020) gas chromatograph (GC)
2. Dual-beam ozone photometer (OZ)
3. Tunable diode laser (TDL) spectrometer for water vapor (WV)

Measurements
Recent Missions
(WB-57 - JSC)
;
(ER-2 - AFRC)
;
(DC-8 - AFRC)
;
(Global Hawk - AFRC)
;
(Gulfstream V - NSF)
;
(Global Hawk - AFRC)
Point(s) of Contact
(Co-I),
(Co-I),
(POC; PI)
Range of Measurement
In situ
Data Delivery
Preliminary data in 1-2 days, final data in approximately 6 months
Weight
49.00 kg
Size
Length
46.00 cm
(L) x
Width
41.00 cm
(W) x
Height
25.00 cm
(H)
Power:
150.00 W
(maximum
300.00 W
)
Has Pressurized Canisters? Yes
Location
Zone 16 on Global Hawk 872.
Upper Q-Bay on ER-2
Main cabin on DC-8
Notes
Different parts of UCATS generate data at different rates:
N2O, SF6 - 70 or 75 seconds
CH4, H2, CO - 140 seconds
CFCs, H1211 - 75 seconds
CHCl3, CCl4 - 150 seconds
Ozone - 2 seconds or 10 seconds
H2O - 1 HZ
TRL
9
Publications