Disclaimer: This material is being kept online for historical purposes. Though accurate at the time of publication, it is no longer being updated. The page may contain broken links or outdated information, and parts may not function in current web browsers. Visit https://espo.nasa.gov for information about our current projects.
Principle: The CU aircraft version of the Aerodyne High-Resolution Time-of-Flight Aerosol Mass Spectrometer (HR-ToF-AMS) detects non-refractory submicron aerosol composition by impaction on a vaporizer at 600°C, followed by electron ionization and time-of-flight mass spectral analysis. Size-resolved composition can be quantified by measuring the arrival times of the aerosol at the vaporizer.
Aircraft Operation: (1 min cycles, can be adjusted to meet mission goals):
46 s total concentration measurements (1 s resolution, can be increased to up to 10 Hz upon request)
5 s speciated size distribution measurements (with improved S/N detection due to ePToF acquisition)
9 s Background + Overhead
Higher accuracy due to flight day calibrations using built-in system
Custom pressure controlled inlet with confirmed performance up to 45 kft
Real Time Data Products:
PM1 Aerosol Mass Concentrations:
Organic aerosol (OA) , SO4, NO3, NH4, Chloride
OA Chemical Markers: f44 (Secondary OA), f57 (hydrocarbon-like OA), f60 (biomass burning OA), f82 (isoprene epoxide-SOA), other fx upon request
More Advanced Products:
- PM1 Seasalt, ClO4, total I, total Br, MSA concentrations
- O/C, H/C, OA/OC, OSc
- Particle organic nitrates (pRONO2)
- Ammonium Balance, estimated pH
- OA components by positive matrix factorization (PMF)
- Particle eddy covariance fluxes of all species
- Speciated Aerosol size distributions
Detection Limits (1s, ng sm-3), (1 min, ng sm-3) from start of the flight (due to custom cryopump):
Sulfate: 40, 15
Nitrate: 15, 6
Ammonium: 3, 1
Chloride: 30, 12
OA: 200, 80
For detailed OA analysis, longer averaging (3-30 s, depending on OA concentration) is needed. A 1 min product is hence provided as well.