Dual-Beam UV-Absorption Ozone Photometer (NOAA O3 Classic)

Status

Operated By: 
PI

The NOAA-O3 instrument consists of a mercury lamp, two sample chambers that can be periodically scrubbed of ozone, and two detectors that measure the 254-nm radiation transmitted through the chamber. The ozone absorption cross-section at this wavelength is accurately known; hence, the ozone number density can be easily calculated. Since the two absorption chambers are identical, virtually continuous measurements of ozone are made by alternating the ambient air sample and ozone scrubbed sample between the two chambers. At a one-second data collection rate, the minimum detectable concentration of ozone (one standard deviation) is 1.5 x 10 10 molecules/cm 3 (0.6ppbv at STP).

Instrument Type: 
Measurements: 
Aircraft: 
Point(s) of Contact: 
Erik Richard (POC; PI), Mike Proffitt (Prev PI)