Disclaimer: This material is being kept online for historical purposes. Though accurate at the time of publication, it is no longer being updated. The page may contain broken links or outdated information, and parts may not function in current web browsers. Visit https://espo.nasa.gov for information about our current projects.

 

NOAA Iodide Ion Time-of-Flight Chemical Ionization Mass Spectrometer (NOAA ToF CIMS)

Status

Operated By: 
PI

Iodide Ion ToF (Time-of-Flight) CIMS (Chemical Ionization Mass Spectrometer)

Principle of the Measurement

Chemical ionization mass spectrometric detection of gas phase organic and inorganic analytes via I- adduct formation

Species Measured

Reactive nitrogen species: HNO3 (nitric acid), HONO (nitrous acid), HO2NO2 (peroxynitric acid), N2O5 (dinitrogen pentoxide), HNCO (isocyanic acid) 
Halogen Species: ClNO2, HCl, HBr, HOBr, HOCl, Cl2, Br2 
Low to intermediate volatility organic species

Time Response

Instrumental response <1 sec, Field response is limited by inlet surface affinity for a particular species

Detection Limit

Precision on 1s data various by species

Accuracy

(15% + 1 pptv) for N2O5
(20% + 1 pptv) for ClNO2
(30% + 15 pptv) for HONO
(25% + 10 pptv) for HO2NO2
(15% + 15 pptv) for HNO3
(20% + 5 pptv) for HNCO
(20% + 15 pptv) for HCOOH
(30% + 1 pptv) for halogenated species

Manufacturer

TOFWERK/Aerodyne Research Inc. (modified)

Instrument Type: 
Measurements: 
Aircraft: 
Point(s) of Contact: 
J. Andy Neuman (POC; PI), Patrick Veres (Co-I)