Disclaimer: This material is being kept online for historical purposes. Though accurate at the time of publication, it is no longer being updated. The page may contain broken links or outdated information, and parts may not function in current web browsers. Visit https://espo.nasa.gov for information about our current projects.
Principle of the Measurement
Chemical ionization mass spectrometric detection of gas phase organic and inorganic analytes via I- adduct formation
Species Measured
Reactive nitrogen species: HNO3 (nitric acid), HONO (nitrous acid), HO2NO2 (peroxynitric acid), N2O5 (dinitrogen pentoxide), HNCO (isocyanic acid)
Halogen Species: ClNO2, HCl, HBr, HOBr, HOCl, Cl2, Br2
Low to intermediate volatility organic species
Time Response
Instrumental response <1 sec, Field response is limited by inlet surface affinity for a particular species
Detection Limit
Precision on 1s data various by species
Accuracy
(15% + 1 pptv) for N2O5
(20% + 1 pptv) for ClNO2
(30% + 15 pptv) for HONO
(25% + 10 pptv) for HO2NO2
(15% + 15 pptv) for HNO3
(20% + 5 pptv) for HNCO
(20% + 15 pptv) for HCOOH
(30% + 1 pptv) for halogenated species
Manufacturer
TOFWERK/Aerodyne Research Inc. (modified)