We report new Fourier transform infrared (FT-IR) spectroscopy measurements of the 17O-enriched isotopologues of CO2 in the m3 region (2200–2450 cm-1, 65–75 THz) with absolute calibration accuracies better than 3 × 10-6 cm-1 (~100 kHz) for strong, isolated transitions. Spectra were obtained using the JPL Bruker IFS-125HR Fourier Transform Spectrometer (JPL-FTS). The JPL-FTS performance was achieved by vigorously maintaining the optical alignment, regularly monitoring its modulation efficiency while minimizing phase errors to deliver absolute frequency calibration and line position accuracies comparable to typical sub-millimeter/THz spectroscopy. Fits of the 00011 00001 fundamental bands for 16O12C17O (627), 17O12C17O (727), and 17O12C18O (728) were obtained with RMS residual values of 2.9 × 10-6 cm-1 for all three isotopologues. These results provide a complete and accurate set of transition frequencies required for remote sensing of planetary atmospheres, especially Venus and Mars, and represent a new level of Doppler measurement accuracies.