PeroxyAcetylNitrate, Aldehydes and Ketones (PANAK)

Status

Operated By: 
PI

Don Hearlth, Rick Kolyer

The Ames PANAK instrument is a computerized 3- channel Capillary Gas Chromatographic system designed for the collection and analysis of low ppt (10-12 v/v) levels of peroxyacyl nitrates (PANs), alkyl nitrates, and tertrachloroethene in Channels 1 and 2; and C2-C3 aldehydes, C1-C2 alcohols, C3-C4 ketones, and C1-C2 nitriles in channel 3. Channels 1 and 2 use ECD detectors and have a sampling frequency of 2.5 minutes. Channel 3 uses a Photo Ionization detector placed in series with a Reduction Gas detector and has a sampling frequency of 5 minutes. The main manifold draws 5 SL/min of ambient air through a heated Teflon lined probe from which each of the three instrument channels draws a 200 ml aliquot of sample air. This aliquot is dried by passing it through a –35 °C cold trap, cooled to -140 °C for constituent pre concentration, and then heat desorbed into the gas chromatographic columns. All calibrations are performed in-flight by using an installed dilution system and in a manner that mimics ambient air sampling. Primary standards are generally referred to a series of permeation tubes. In addition high concentration standards are also carried on board. Sensitivities under typical conditions are: 1-3 ppt PANs, 1-5 ppt alkyl nitrates, 5-20 ppt OVOC, and 20-30 ppt nitriles.

Instrument Type: 
Aircraft: 
Point(s) of Contact: 
Hanwant Singh (POC; PI)