A new non-resonant laser-induced fluorescence instrument for the airborne in situ measurement of formaldehyde

St. Clair, J.M., A.K. Swanson, S.A. Bailey, G.M. Wolfe, J. Marrero, L.T. Iraci, J.G. Hagopian, and T.F. Hanisco (2017), A new non-resonant laser-induced fluorescence instrument for the airborne in situ measurement of formaldehyde, Atmos. Meas. Tech., 10, 4833-4844, doi:10.5194/amt-10-4833-2017.
Abstract

A new in situ instrument for gas-phase formaldehyde (HCHO), COmpact Formaldehyde FluorescencE Experiment (COFFEE), is presented. COFFEE utilizes nonresonant laser-induced fluorescence (NR-LIF) to measure HCHO, with 300 mW of 40 kHz 355 nm laser output exciting multiple HCHO absorption features. The resulting HCHO fluorescence is collected at 5 ns resolution, and the fluorescence time profile is fit to yield the ambient HCHO mixing ratio. Typical 1σ precision at ∼ 0 pptv HCHO is 150 pptv for 1 s data. The compact instrument was designed to operate with minimal in-flight operator interaction and infrequent maintenance (1–2 times per year). COFFEE fits in the wing pod of the Alpha Jet stationed at the NASA Ames Research Center and has successfully collected HCHO data on 27 flights through 2017 March. The frequent flights, combined with a potentially long-term data set, makes the Alpha Jet a promising platform for validation of satellite-based column HCHO.

PDF of Publication
Download from publisher's website
Research Program
Upper Atmosphere Research Program (UARP)
Mission
AJAX