Disclaimer: This material is being kept online for historical purposes. Though accurate at the time of publication, it is no longer being updated. The page may contain broken links or outdated information, and parts may not function in current web browsers. Visit https://espo.nasa.gov for information about our current projects.

 

Synonyms: 
NSF G-V
G-V NCAR
G-5 NCAR
G-5 NSF
G-V NSF

Closed-path Laser Hygrometer

The University of Colorado closed-path tunable diode laser hygrometer (CLH) is based on the water vapor hygrometers designed by R. D. May (Maycomm, Inc.). CLH is coupled to a heated, forward-facing inlet that enhances particulate water by anisokinetic sampling. Ice water content (IWC) is derived from the measurement of enhanced total water, with knowledge of the instrument sampling characteristics, particle size distributions and ambient water vapor.

In contrast to the open-path systems of similar heritage, the CLH, which was designed for operation in the troposphere on commercial aircraft, has a single-pass absorption cell (27.62 cm long). The light source is a room-temperature solid-state laser that puts out 3-5 mW of radiation at 1.37 mm (7306.752 cm-1).

Instrument Type: 
Measurements: 
Aircraft: 
Gulfstream V - NSF, WB-57 - JSC, DC-8- AFRC
Point(s) of Contact: 

Focused Cavity Aerosol Spectrometer

The FCAS II sizes particles in the approximate diameter range from 0.07 mm to 1 mm. Particles are sampled from the free stream with a near isokinetic sampler and are transported to the instrument. They are then passed through a laser beam and the light scattered by individual particles is measured. Particle size is related to the scattered light. The data reduction for the FCAS II takes into account the water which is evaporated from the particle in sampling and the effects of anisokinetic sampling (Jonsson et al., 1995).

The FCAS II and its predecessors have provided accurate aerosol size distribution measurements throughout the evolution of the volcanic cloud produced by the eruption of Mt. Pinatubo. (Wilson et al., 1993). Near co-incidences between FCAS II and SAGE II measurements show good agreement between optical extinctions calculated from FCAS size distributions and extinctions measured by SAGE II.

Accuracy: The instrument has been calibrated with monodisperse aerosol carrying a single charge. The FCAS III and the electrometer agree to within 10%. Sampling errors may increase the uncertainty but a variety of comparisons suggests that total uncertainties in aerosol surface are near 30% (Jonsson, et al., 1995).

Precision: The precision equals 1/ÖN where N is the number of particles counted. In many instances the precision on concentration measurements may reach 7% for 0.1 Hz data. If better precision is desired, it is necessary only to accumulate over longer time intervals.

Response Time: Data are processed at 0.1 Hz. However, the response time depends upon the precision required to detect the change in question. Small changes may require longer times to detect. Plume measurements may be processed with 1 s resolution.

Weight: Approximately 50 lbs.

Instrument Type: 
Aircraft: 
Point(s) of Contact: 