Disclaimer: This material is being kept online for historical purposes. Though accurate at the time of publication, it is no longer being updated. The page may contain broken links or outdated information, and parts may not function in current web browsers. Visit https://espo.nasa.gov for information about our current projects.

 

Synonyms: 
WB-57
WB57
Associated content: 

Harvard Halogen Instrument

Many changes from the original Chlorine Nitrate instrument. NO2 instrument removed. New inlet with orifice for one halogen duct, addition of vacuum scroll pump, new RF oscillators and amplifiers, new RF frequency, new lamp housings and cooling for lamp modules. Flew in MACPEX without dissociation heaters, i.e., focus on BrO and ClO measurements and not measure ClONO2 or ClOOCl.

Instrument Type: 
Measurements: 
Aircraft: 
Point(s) of Contact: 

Advanced Whole Air Sampler

50 samples/flight

New control system

Fill times
–14 km 30 – 40 sec
–16 km 40 – 50 sec
–18 km 50 – 60 sec
–20 km 100 – 120 sec (estimated)

Analysis in UM lab: GC/MS; GC/FID; GC/ECD

Instrument Type: 
Measurements: 
Point(s) of Contact: 

Quantum Cascade Laser System

The Harvard QCLS (DUAL and CO2) instrument package contains 2 separate optical assemblies and calibration systems, and a common data system and power supply. The two systems are mounted in a single standard HIAPER rack, and are described separately below:

The Harvard QCL DUAL instrument simultaneously measures CO, CH4, and N2O concentrations in situ using two thermoelectrically cooled pulsed-quantum cascade lasers (QCL) light sources, a multiple pass absorption cell, and two liquid nitrogen-cooled solid-state detectors. These components are mounted on a temperature-stabilized, vibrationally isolated optical bench with heated cover. The sample air is preconditioned using a Nafion drier (to remove water vapor), and is reduced in pressure to 60 mbar using a Teflon diaphragm pump. The trace gas mixing ratios of air flowing through the multiple pass absorption cell are determined by measuring absorption from their infrared transition lines at 4.59 microns for CO and 7.87 microns for CH4 and N2O using molecular line parameters from the HITRAN data base. In-flight calibrations are performed by replacing the air sample with reference gas every 10 minutes, with a low-span and a high-span gas every 20 minutes. A prototype of this instrument was flown on the NOAA P3 in the summer of 2004.

The Harvard QCL CO2 instrument measures CO2 concentrations in situ using a thermoelectrically cooled pulsed-quantum cascade laser (QCL) light source, gas cells, and liquid nitrogen cooled solid-state detectors. These components are stabilized along the detection axis, vibrationally isolated, and housed in a temperature-controlled pressure vessel. Sample air enters a rear-facing inlet, is preconditioned using a Nafion drier (to remove water vapor), then is reduced in pressure to 60 mbar using a Teflon diaphragm pump. A second water trap, using dry ice, reduces the sample air dewpoint to less than –70C prior to detection. The CO2 mixing ratio of air flowing through the sample gas cell is determined by measuring absorption from a single infrared transition line at 4.32 microns relative to a reference gas of known concentration. In-flight calibrations are performed by replacing the air sample with reference gas every 10 minutes, and with a low-span and a high-span gas every 20 minutes.

Instrument Type: 
Measurements: 
Point(s) of Contact: 

NCAR NOxyO3

The NCAR NOxyO3 instrument is a 4-channel chemiluminescence instrument for the measurement of NO, NO2, NOy, and O3. NOx (NO and NO2) is critical to fast chemical processes controlling radical chemistry and O3 production. Total reactive nitrogen (NOy = NO + NO2 + HNO3 + PANs + other organic nitrates + HO2NO2 + HONO + NO3 + 2*N2O5 + particulate NO3- + …) is a useful tracer for characterizing air masses since it has a tendency to be conserved during airmass aging, as NOx is oxidized to other NOy species.

NOx (NO and NO2), NOy (total reactive nitrogen), and O3 are measured using the NCAR 4-channel chemiluminescence instrument, previously flown on the NASA WB-57F and the NCAR C130. NO is measured via addition of reagent O3 to the sample flow to generate the chemiluminescent reaction producing excited NO2, which is detected by photon counting with a dry-ice cooled photomultiplier tube. NO2 is measured as NO following photolytic conversion of NO2, with a time response of about 3 sec due to the residence time in the photolysis cell. NO is measured with an identical time response due to use of a matching volume. NOy is measured via Au-catalyzed conversion of reactive nitrogen species to NO, in the presence of CO, with a time response of slightly better than 1 sec. O3 is measured using the same chemiluminescent reaction but with the addition of reagent NO to the sample flow. Time response for the ozone measurement is slightly better than 1 s.

Instrument Type: 
Measurements: 
Point(s) of Contact: 