Disclaimer: This material is being kept online for historical purposes. Though accurate at the time of publication, it is no longer being updated. The page may contain broken links or outdated information, and parts may not function in current web browsers. Visit https://espo.nasa.gov for information about our current projects.

 

Synonyms: 
CRAVE
CR AVE
Associated content: 

Counterflow Virtual Impactor

The NCAR counterflow virtual impactor (CVI) (Noone et al., 1988; Twohy et al., 1997) is an airborne instrument that can be used for studies of aerosol/cloud interactions, cloud physics, and climate. At the CVI inlet tip, cloud droplets or ice crystals larger than about 8 µm aerodynamic diameter are separated from the interstitial aerosol and impacted into dry nitrogen gas. This separation is possible via a counterflow stream of nitrogen out the CVI tip, which assures that only larger particles (cloud droplets or ice crystals) are sampled. Because droplets or crystals in a sampling volume of about 200 l/min are impacted into a sample stream of approximately 10 l/min, concentrations within the CVI are significantly enhanced. The water vapor and non-volatile residual nuclei remaining after droplet evaporation are sampled downstream of the inlet with selected instruments. These may include a Lyman-alpha or similar hygrometer, a condensation nucleus counter, an optical particle counter, filters for chemical analyses, or user instruments.

Point(s) of Contact: 

Focused Cavity Aerosol Spectrometer

The FCAS II sizes particles in the approximate diameter range from 0.07 mm to 1 mm. Particles are sampled from the free stream with a near isokinetic sampler and are transported to the instrument. They are then passed through a laser beam and the light scattered by individual particles is measured. Particle size is related to the scattered light. The data reduction for the FCAS II takes into account the water which is evaporated from the particle in sampling and the effects of anisokinetic sampling (Jonsson et al., 1995).

The FCAS II and its predecessors have provided accurate aerosol size distribution measurements throughout the evolution of the volcanic cloud produced by the eruption of Mt. Pinatubo. (Wilson et al., 1993). Near co-incidences between FCAS II and SAGE II measurements show good agreement between optical extinctions calculated from FCAS size distributions and extinctions measured by SAGE II.

Accuracy: The instrument has been calibrated with monodisperse aerosol carrying a single charge. The FCAS III and the electrometer agree to within 10%. Sampling errors may increase the uncertainty but a variety of comparisons suggests that total uncertainties in aerosol surface are near 30% (Jonsson, et al., 1995).

Precision: The precision equals 1/ÖN where N is the number of particles counted. In many instances the precision on concentration measurements may reach 7% for 0.1 Hz data. If better precision is desired, it is necessary only to accumulate over longer time intervals.

Response Time: Data are processed at 0.1 Hz. However, the response time depends upon the precision required to detect the change in question. Small changes may require longer times to detect. Plume measurements may be processed with 1 s resolution.

Weight: Approximately 50 lbs.

Instrument Type: 
Aircraft: 
Point(s) of Contact: 

Condensation Nuclei Counter

The CNC counts particles in the approximate diameter range from 0.006 m to 2 m. The instrument operates by exposing the articles to saturated Flourinert vapor at 28 C and then cooling the sample in a condenser at 5 C. The supersaturation of the vapor increases as it is cooled and the vapor condenses on the particles causing them to grow to sizes which are easily detected. The resulting droplets are passed through a laser beam and the scattered light is detected. Individual particles are counted and are referred to as condensation nuclei (CN). Two CN Counters are provided in the instrument. One counts the particles after sampling from the atmosphere and the second counts particles that have survived heating to 192C. Lab experiments show that pure sulfuric acid particles smaller than 0.05 mm are volatilized in the heater. The heated channel detects when small particles are volatile and permits speculation about the composition. The CNC II contains an impactor collector which permits the collection of particles on electron microscope grids for later analysis. The collector consists of a two stages. In the first stage the pressure of the sample is reduced by a factor of two without loosing particles by impaction on walls. The second stage consists of a thin plate impactor which collect efficiently even at small Reynolds numbers. The system collects particles as small as 0.02 m at WB-57 cruise altitudes. As many as 25 samples can be collected in a flight.

Instrument Type: 
Measurements: 
Point(s) of Contact: 

GeoCAPE Airborne Spectrometer

Two spectrographs + HD video camera

Air Quality (AQ) 304:520 nm 0.8 nm resolution (NO2, O3, UV absorbing aerosols, SO2, HCHO)

Ocean Color (OC) 460:900 nm 1.5 nm resolution

Video camera (2592x1936 pixels) –3 pixel FWHM

Instrument Type: 
Measurements: 
Point(s) of Contact: 

Charged-coupled device Actinic Flux Spectroradiometers

The Charged-coupled device Actinic Flux Spectroradiometers (CAFS) instruments measure in situ down- and up-welling radiation and combine to provide 4 pi steradian actinic flux density spectra from 280 to 650 nm. The sampling resolution is ~0.8 nm with a full width at half maximum (FWHM) of 1.7 nm at 297 nm. From the measured flux, photolysis frequencies are calculated for ~40 important atmospheric trace gases including O3, NO2, HCHO, HONO and NO3 using a modified version of the NCAR Tropospheric Ultraviolet and Visible (TUV) radiative transfer model. The absolute spectral sensitivity of the instruments is determined in the laboratory with 1000 W NIST-traceable tungsten-halogen lamps with a wavelength dependent uncertainty of 3–5%. During deployments, spectral sensitivity is assessed with secondary calibration lamps while wavelength assignment is tracked with Hg line sources and comparisons to spectral features in the extraterrestrial flux. The optical collectors are characterized for angular and azimuthal response and the effective planar receptor distance. CAFS have an excellent legacy of performance on the NASA DC-8 and WB-57 platforms during atmospheric chemistry and satellite validation mission. These include AVE Houston 2004 and 2005, PAVE, CR-AVE, TC4, ARCTAS, DC3, SEAC4RS, KORUS-AQ, ATom and FIREX-AQ. For FIREX-AQ, upgraded electronics and cooling reduced noise and allowed for a decrease to 1 Hz acquisition.

Instrument Type: 
Point(s) of Contact: 

Argus Tunable Diode Laser Instrument

Argus is a two channel, tunable diode laser instrument set up for the simultaneous, in situ measurement of CO (carbon monoxide), N2O (nitrous oxide) and CH4 (methane) in the troposphere and lower stratosphere. The instrument measures 40 x 30 x 30 cm and weighs 21 kg. An auxiliary, in-flight calibration system has dimensions 42 x 26 x 34 cm and weighs 17 kg.

The instrument is an absorption spectrometer operating in rapid scan, secondharmonic mode using frequency-modulated tunable lead-salt diode lasers emitting in the mid-infrared. Spectra are co-added for two seconds and are stored on a solid state disk for later analysis. The diode laser infrared beam is shaped by two anti-refection coated lenses into an f/40 beam focused at the entrance aperture of a multi-pass Herriott cell. The Herriott cell is common to both optical channels and is a modified astigmatic cell (New Focus Inc., Santa Clara, California).

The aspherical mirrors are coated with protected silver for optimal infrared reflectivity. The cell is set up for a 182-pass state for a total path of 36m. The pass number can be confirmed by visual spot pattern verification on the mirrors observed through the glass cell body when the cell is illuminated with a visible laser beam. However, instrument calibration is always carried out using calibrated gas standards with the Argus instrument operating at its infrared design wavelengths, 3.3 and 4.7 micrometers respectively for CH4 and CO detection. The electronic processing of the second harmonic spectra is done by standard phase sensitive amplifier techniques with demodulation occurring at twice the laser modulation frequency of 40 kHz. To optimize the secondharmonic signal amplitude in a changing ambient pressure environment the laser modulation amplitude is updated every 2 seconds to its optimal theoretical value based upon the measured pressure in the Herriott cell.

Measurements: 
Aircraft: 
Point(s) of Contact: 

Chemical Ionization Mass Spectrometer

The NOAA chemical ionization mass spectrometer (CIMS) instrument was developed for high-precision measurements of gaseous nitric acid (HNO3) specifically under high- and variable-humidity conditions in the boundary layer. The instrument’s background signals (i.e., signals detected when HNO3-free air is measured), which depend on the humidity and HNO3 concentration of the sample air, are the most important factor affecting the limit of detection (LOD). A new system to provide HNO3-free air without changing both the humidity and the pressure of the sampled air was developed to measure the background level accurately. The detection limit was about 23 parts per trillion by volume (pptv) for 50-s averages. Field tests, including an intercomparison with the diffusion scrubber technique, were carried out at a surface site in Tokyo, Japan, in October 2003 and June 2004. A comparison between the measured concentrations of HNO3 and particulate nitrate indicated that the interference from particulate nitrate was not detectable (i.e., less than about 1%). The intercomparison indicated that the two independent measurements of HNO3 agreed to within the combined uncertainties of these measurements.

Instrument Type: 
Measurements: 
Point(s) of Contact: 

Cloud, Aerosol, and Precipitation Spectrometer

The CAPS is a combination probe designed around the newest technologies and the experience gained with over 20 years of using similar probes. It meets the goals of measuring a large range of particle sizes--0.5μm to 1.55mm--with one probe, thus minimizing space, cable connections, and data systems necessary for measurement of this range. Today's technology also provides the CAPS the processing power necessary to perform at speeds up to 200m/s. An intuitive graphical user interface, the Particle Analysis and Collection System (PACS), at the host computer, provides simple but powerful control of measurement parameters, while simultaneously displaying on-the-fly size distributions and derived parameters. All data interfaces are done via line drivers meeting the RS-422 electrical specification, allowing cable lengths of up to 100 meters--an improvement over RS-232 lines capable of only 15-meter cable lengths.

Instrument Type: 
Aircraft: 
Point(s) of Contact: 

2D-S Stereo Probe

The 2D-S Stereo Probe is an optical imaging instrument that obtains stereo cloud particle images and concentrations using linear array shadowing. Two diode laser beams cross at right angles and illuminate two linear 128-photodiode arrays. The lasers are single-mode, temperature-stabilized, fiber-coupled diode lasers operating at 45 mW. The optical paths are arbitrarily labeled the “vertical” and “horizontal” probe channels, but the verticality of each channel actually depends on how the probe is oriented on an aircraft. The imaging optical system is based on a Keplerian telescope design having a (theoretical) primary system magnification of 5X, which results in a theoretical effective size of (42.5 µm + 15 µm)/5 = 11.5 µm. However, actual lenses and arrays have tolerances, so it is preferable to measure the actual effective pixel size by dropping several thousands of glass beads with known diameters through the object plane of the optics system.

Instrument Type: 
Point(s) of Contact: 

Aircraft Laser Infrared Absorption Spectrometer

ALIAS (Aircraft Laser Infrared Absorption Spectrometer) measures total water, total water isotopes, carbon monoxide, and carbon dioxide isotope ratios. No other instrument provides real-time measurements of carbon dioxide isotope ratios which are clear identifiers of atmospheric transport (18O/17O/16O for stratospheric intrusion, 13C/12C for anthropogenic signals). ALIAS easily adapts to changing mission priorities and can be configured to measure HCl, CH4, SO2, and N2O by simply replacing a semiconductor laser. These measurements contribute to Atmospheric Composition Focus Area research by providing key data on how convective processes affect stratospheric composition, the development of cirrus particles and their affect on Earth's radiative balance, and health of the ozone layer through measurement of chlorine partitioning.

Measurements: 
Aircraft: 
Point(s) of Contact: 