Disclaimer: This material is being kept online for historical purposes. Though accurate at the time of publication, it is no longer being updated. The page may contain broken links or outdated information, and parts may not function in current web browsers. Visit espo.nasa.gov for information about our current projects.

 

PAN and Trace Hydrohalocarbon ExpeRiment (PANTHER)

Status

Operated By: 
PI

PANTHER uses Electron Capture Detection and Gas Chromatography (ECD-GC) and Mass Selective Detection and Gas Chromatography (MSD-GC) to measure numerous trace gases, including Methyl halides, HCFCs, PAN, N20, SF6, CFC-12, CFC-11, Halon-1211, methyl chloroform, carbon tetrachloride.

3 ECD (electron capture detectors), packed columns (OV-101, Porpak-Q, molecular sieve).

1 ECD with a TE (thermal electric) cooled RTX-200 capillary column.

2-channel MSD (mass selective detector). The MSD analyses two independent samples concentrated onto TE cooled Haysep traps, then passed through two temperature programmed RTX-624 capillary columns.

With the exception of PAN, all channels of chromatography are normalized to a stable in-flight calibration gas references to NOAA scales. The PAN data is normalized to an in-flight PAN source of ≈ 100 ppt with ±5 % reproducibility. This source is generated by efficient photolytic conversion of NO in the presence of acetone. Detector non-linearity is taken out by lab calibrations for all molecules.